TITLE

PERFORMANCE OF THE ELECTRONIC PERSONAL DOSEMETER FOR NEUTRON `SAPHYDOSE-N' AT DIFFERENT WORKPLACES OF NUCLEAR FACILITIES

AUTHOR(S)
Lahaye, T.; Chaul, Q.; Ménard, S.; Lacoste, V.; Muller, H.; Luszik-Bhadra, M.; Reginatto, M.; Bruguier, P.
PUB. DATE
September 2006
SOURCE
Radiation Protection Dosimetry;2006, Vol. 120 Issue 1-4, p383
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
This paper mainly aims at presenting the measurements and the results obtained with the electronic personal neutron dosemeter Saphydose-N at different facilities. Three campaigns were led in the frame of the European contract EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields'). The first one consisted in the measurements at the IRSN French research laboratory in reference neutron fields generated by a thermal facility (SIGMA), radionuclide ISO sources (241AmBe; 252Cf; 252Cf(D2O)\Cd) and a realistic spectrum (CANEL/T400). The second one was performed at the Krümmel Nuclear Power Plant (Germany) close to the boiling water reactor and to a spent fuel transport cask. The third one was realised at Mol (Belgium), at the VENUS Research Reactor and at Belgonucléaire, a fuel processing factory.